Impact of different IMRT techniques to improve conformity and normal tissue sparing in upper esophageal cancer

Amin E Amin, Mohamed Kelaney, Samah K Elshamndy, Osiris W. Guirguis

Abstract


Purpose: Intensity modulated radiotherapy (IMRT) for cervical esophageal cancer is challenging. Although IMRT techniques using inverse planning algorithms are facilitating the treatment planning process, the irradiation dose to the normal tissues can be a critical issue. This study was performed to investigate the effect of beam numbers and their directions and local optimization on: (1) dose conformity and homogeneity to the planning target volume (PTV) and (2) dose to the organ at risks (OARs).

Methods: Four upper esophageal cancer cases were randomly selected for this treatment planning study. Eight IMRT plans were generated for each case with the same dose-volume constraints but with different beam numbers and arrangements. Local optimization using regular structures drawn automatically around the PTV with margins from 0.5-1.5 cm was performed. IMRT plans were evaluated with respect to isodose distributions, dose-volume histograms (DVHs) parameters, homogeneity index (HI), and conformity index (CI). The statistical comparison between the types of plans was done using the One Way ANOVA test.

Results: The results showed that IMRT using three or five beams was not sufficient to obtain good dose optimization. The seven field plans showed the best coverage for the PTV with tolerable doses for the OARs, and the beam orientation was very critical. Increasing beams (Bs) number from 7 to 13 did not show significant differences in the PTV coverage, while the mean lung dose was increased. The PTV coverage were 95.1, 95.1, 98.1, 97.3, 97.3, 97.3, 97.0, and 97.0% for 3Bs, 5Bs, 7Bs, 9Bs, 13Bs, 7Bs(30), 7Bs(60) (beam angles were changed from 0o to 30o and 60o), and 7Bs(R) (seven IMRT plans with ring), respectively. The mean heart dose did not exceed 0.36 Gy with p < 0.05. For lung doses, the best plan was the one with 9Bs which reduced lung volume doses V20Gy (%) and V30Gy (%), and reduced mean lung dose from 5.4 to 4.5 Gy with p < 0.05 for 7Bs(R) plans. IMRT improved the homogeneity indices (p > 0.05), yet conformity was better with 9Bs and 7Bs(R) IMRT plans with p < 0.05.

Conclusion: Seven equispaced coplanar intensity-modulated beams with an addition of a ring structure can produce desirable dose distributions to the PTV. Moreover, dose-volume of exposed normal lung can be reduced with 9Bs and 7Bs(R) IMRT plans.


Keywords


Intensity Modulated Radiotherapy; Esophageal Cancer; Local Optimization; Planning Target Volume; Organ at Risks

Full Text:

PDF

References


Burmeister BH, Dickie G, Smithers BM, Hodge R, Morton K. Thirty-four patients with carcinoma of the cervical esophagus treated with chemoradiation therapy. Arch Otolaryngol Head Neck Surg 2000; 126:205-8.

Huang SH, Lockwood G, Brierley J, et al. Effect of concurrent high-dose cisplatin chemotherapy and conformal radiotherapy on cervical esophageal cancer survival. Int J Radiat Oncol Biol Phys 2008; 71:735-40.

Yamada K, Murakami M, Okamoto Y, et al. Treatment results of radiotherapy for carcinoma of the cervical esophagus. Acta Oncol 2006; 45:1120-5.

Tai P, Van Dyk J, Yu E, et al. Variability of target volume delineation in cervical esophageal cancer. Int J Radiat Oncol Biol Phys 1998; 42:277-88.

Tai P, Van Dyk J, Yu E, et al. Radiation treatment for cervical esophagus: patterns of practice study in Canada, 1996. Int J Radiat Oncol Biol Phys 2000; 47:703-12.

Wu Q, Manning M, Schmidt-Ullrich R, Mohan R. The potential for sparing of parotids and escalation of biologically effective dose with intensity-modulated radiation treatments of head and neck cancers: a treatment design study. Int J Radiat Oncol Biol Phys 2000; 46:195-205.

Wu VW, Kwong DL, Sham JS. Target dose conformity in 3-dimensional conformal radiotherapy and intensity modulated radiotherapy. Radiother Oncol 2004; 71:201-6.

Hsiung CY, Yorke ED, Chui CS, et al. Intensity-modulated radiotherapy versus conventional three-dimensional conformal radiotherapy for boost or salvage treatment of nasopharyngeal carcinoma. Int J Radiat Oncol Biol Phys 2002; 53:638-47.

Nutting CM, Bedford JL, Cosgrove VP, et al. A comparison of conformal and intensity-modulated techniques for oesophageal radiotherapy. Radiother Oncol 2001; 61:157-63.

Nutting C, Dearnaley DP, Webb S. Intensity modulated radiation therapy: a clinical review. Br J Radiol 2000; 73:459-69.

Lee N, Xia P, Fischbein NJ, et al. Intensity-modulated radiation therapy for head-and-neck cancer: the UCSF experience focusing on target volume delineation. Int J Radiat Oncol Biol Phys 2003; 57:49-60.

Fu WH, Wang LH, Zhou ZM, et al. Comparison of conformal and intensity-modulated techniques for simultaneous integrated boost radiotherapy of upper esophageal carcinoma. World J Gastroenterol 2004; 10:1098-102.

Wu V, Sham J, Kwong D. Inverse planning in three-dimensional conformal and intensity-modulated radiotherapy of mid-thoracic esophageal cancer. Br J Radiol 2004; 77:568-72.

Spirou SV, Chui CS. Generation of arbitrary intensity profiles by dynamic jaws or multileaf collimators. Med Phys 1994; 21:1031-41.

Nutting CM, Bedford JL, Cosgrove VP, et al. Intensity-modulated radiotherapy reduces lung irradiation in patients with carcinoma of the oesophagus. Front Radiat Ther Oncol 2002; 37:128-31.

Chandra A, Guerrero TM, Liu HH, et al. Feasibility of using intensity-modulated radiotherapy to improve lung sparing in treatment planning for distal esophageal cancer. Radiother Oncol 2005; 77:247-53.

Amankwaa-Frempong E, Vernimmen F, Blay S, Ezhilalan R. Irradiation of lung and esophagus tumors: A comparison of dose distributions calculated by anisotropic analytical algorithm and pencil beam convolution algorithm, a retrospective dosimetric study. Int J Cancer Ther Oncol 2014; 2:020210.

Chaikh A, Giraud J, Balosso J. A method to quantify and assess the dosimetric and clinical impact resulting from the heterogeneity correction in radiotherapy for lung cancer. Int J Cancer Ther Oncol 2014; 2:020110.

Rana SB. Dose prediction accuracy of anisotropic analytical algorithm and pencil beam convolution algorithm beyond high density heterogeneity interface. South Asian J Cancer 2013; 2:26-30.

Ojala J. The accuracy of the Acuros XB algorithm in external beam radiotherapy: a comprehensive review. Int J Cancer Ther Oncol 2014; 2:020417.

Rana S, Rogers K. Radiobiological evaluation of dose calculation algorithms in Rapid Arc planning of esophageal cancer treatment plans. J Solid Tumors 2013; 3:44-52.




DOI: http://dx.doi.org/10.14319/ijcto.0301.13

Creative Commons License
This work is licensed under a Creative Commons Attribution 3.0 License.

------------------------------------------------------------

International Journal of Cancer Therapy and Oncology (ISSN 2330-4049)

© International Journal of Cancer Therapy and Oncology (IJCTO)

To make sure that you can receive messages from us, please add the 'ijcto.org' domain to your e-mail 'safe list'. If you do not receive e-mail in your 'inbox', check your 'bulk mail' or 'junk mail' folders.

------------------------------------------------------------

Number of visits since October, 2013
AmazingCounters.com