Total Body Irradiation using VMAT (RapidArc): A Planning Study of a novel treatment delivery method

Santam Chakraborty, Suja Cheruliyil, Resmi Bharathan, Geetha Muttath

Abstract


Purpose: To evaluate the feasibility of using volumetric modulated arc therapy (VMAT) using RapidArc to deliver total body irradiation (TBI) treatment.

Methods: VMAT planning was performed a whole body computed tomography (CT) data set using Rapid Arc. The planning target volumes included entire body trimmed to 3 mm below the skin. The organs at risk included the lungs and kidneys. A dose of 12 Gy in 10 fractions was prescribed to the target volume. The VMAT-TBI technique consisted of three isocentres and three overlapping arcs: the head and neck, the chest, and the pelvis. The plans were prescribed to ensure, at a minimum, 95% planning target volume dose coverage with the prescription dose (percentage of volume receiving dose of 12 Gy was 95%) and maximum dose of 109.8%. Mean dose to lung was restricted at 8.6Gy.

Results: The total body volume in the study was 15469cm3 and the PTV volume was 11322cm3. The mean dose to PTV was 104%. The homogeneity index was 0.09. Sparing of normal tissues with adequate coverage of skeletal bones was shown to be feasible with Rapid Arc. The study demonstrates that VMAT is feasible for TBI treatment. Unlike conventional TBI chest wall boost with electrons was not required.

Conclusion: The technique for total body irradiation using RapidArc VMAT was found feasible and is undergoing further studies prior to clinical use.


Keywords


Total Body Radiation; Volumetric Modulated Arc Therapy; RapidArc; Multiple Isocenters

Full Text:

PDF

References


Van Dyk J, Galvin JM, Glasgow GP, et al. The physical aspects of total and half body photon irradiation: a report of Task Group 29, Radiation Therapy Committee Association of Physicists in Medicine. American Association of Physicists in Medicine 1986.

Quast U. Whole body radiotherapy: A TBI-guideline. J Med Phys 2006; 31:5-12.

Miralbell R, Rouzaud M, Grob E, et al. Can a total body irradiation technique be fast and reproducible? Int J RadiatOncolBiolPhys 1994; 29:1167-73.

Chen HH, Wu J, Chuang KS, et al. Total body irradiation with step translation and dynamic field matching. Biomed Res Int 2013; 2013:216034.

Hui SK, Kapatoes J, Fowler J, et al. Feasibility study of helical tomotherapy for total body or total marrow irradiation. Med Phys 2005; 32:3214-24.

Zhuang AH, Liu A, Schultheiss TE, Wong JY. Dosimetric study and verification of total body irradiation using helical tomotherapy and its comparison to extended SSD technique. Med Dosim 2010; 35:243-9.

Teoh M, Clark CH, Wood K, et al. Volumetric modulated arc therapy: a review of current literature and clinical use in practice. Br J Radiol 2011; 84:967-96.

Aydogan B, Yeginer M, Kavak GO, et al. Total marrow irradiation with RapidArc volumetric arc therapy. Int J RadiatOncolBiolPhys 2011; 81:592-9.

Mancosu P, Navarria P, Castagna L, et al. Anatomy driven optimization strategy for total marrow irradiation with a volumetric modulated arc therapy technique. J ApplClin Med Phys 2012; 13:3653.

Surucu M, Yeginer M, Kavak GO, et al. Verification of dose distribution for volumetric modulated arc therapy total marrow irradiation in a humanlike phantom.Med Phys 2012; 39:281-8.

Han C, Schultheisss TE, Wong JYC. Dosimetric study of volumetric modulated arc therapy fields for total marrow irradiation. RadiotherOncol 2012; 102:315-20.

Zhuang T, Wu Q. Generating arbitrary one-dimensional dose profiles using rotational therapy. Phys Med Biol 2010; 55:6263-77.

Low DA, Harms WB, Mutic S, Purdy JA. A technique for the quantitative evaluation of dose distributions. Med Phys 1998; 25:656-61.

Wolden SL, Rabinovitch RA, Bittner NHJ, et al. American College of Radiology (ACR) and American Society for Radiation Oncology (ASTRO) practice guideline for the performance of total body irradiation (TBI). Am J ClinOncol 2013; 36:97-101.

Peñagarícano JA, Chao M, Van Rhee F, et al. Clinical feasibility of TBI with helical tomotherapy. Bone Marrow Transplant 2011; 46:929-35.

Schultheiss TE, Wong J, Liu A, et al. Image-guided total marrow and total lymphatic irradiation using helical tomotherapy. Int J RadiatOncolBiolPhys 2007; 67:1259-67.

Chandraraj V, Stathakis S, Manickam R, et al. Comparison of four commercial devices for RapidArc and sliding window IMRT QA. J ApplClin Med Phys 2011; 12:3367.

Oinam AS, Singh L. Verification of IMRT dose calculations using AAA and PBC algorithms in dose buildup regions. J ApplClin Med Phys 2010; 11:3351.

Evans MDC, Ruo R, Seuntjens J, et al. Poster-Thur Eve-31: RapidArc total body photon irradiation: A feasibility study. Med Phys 2012; 39:4630.




DOI: http://dx.doi.org/10.14319/ijcto.0302.8

Creative Commons License
This work is licensed under a Creative Commons Attribution 3.0 License.

------------------------------------------------------------

International Journal of Cancer Therapy and Oncology (ISSN 2330-4049)

© International Journal of Cancer Therapy and Oncology (IJCTO)

To make sure that you can receive messages from us, please add the 'ijcto.org' domain to your e-mail 'safe list'. If you do not receive e-mail in your 'inbox', check your 'bulk mail' or 'junk mail' folders.

------------------------------------------------------------

Number of visits since October, 2013
AmazingCounters.com