Statistical methods to evaluate the correlation between measured and calculated dose using quality assurance method in IMRT

Abdulhamid Chaikh, Charline Desgranges, Jacques Balosso

Abstract


Purpose: the objective of this study is to validate a procedure based on a statistical method to assess the agreement and the correlation between measured and calculated dose in the process of quality assurance (QA) for Intensity-Modulated Radiation Therapy (IMRT).

Patients and methods: 10 patients including 56 fields for head and neck cancer treatment were analyzed. For each patient, one treatment plan was generated using Eclipse TPS®. To compare the calculated dose with measured dose a CT-scan of solid water slabs (30 × 30 × 15 cm3) was used. The measurements were done for absolute dose by a pinpoint ionization chamber and 2D dose distributions using electronic portal imaging device dosimetry. Six criteria levels were applied for each case (3%, 3 mm), (4%, 3 mm), (5%, 3 mm), (4%, 4 mm), (5%, 4 mm) and (5%, 5 mm). The normality of the data and the variance homogeneity were tested using Shapiro-Wilks test and Levene’s test, respectively. Wilcoxon signed-rank paired test was used to calculate p-value. Bland-Altman method was used to calculate the limit of agreement between calculated and measured doses and to draw a scatter plot. The correlation between calculated and measured doses was assessed using Spearman’s rank test.

Results: The statistical tests indicate that the data do not fulfill normal distribution, p < 0.001 and had a homogenous variance, p = 0.85. The upper and lower limit of agreements for absolute dose measurements were 6.44% and -6.40%, respectively. Wilcoxon test indicated a significance difference between calculated and measured dose with ionization chamber, p = 0.01. Spearman’s test indicated a strong correlation between calculated and absolute measured dose, ρ = 0.99. Therefore, there is a lack of correlation between dose difference for absolute dose measurements and gamma passing rates for 2D dose measurements.

Conclusion: the statistical tests showed that the common acceptance criteria’s using gamma evaluation are not able to predict the dose difference for a global treatment plan or per beam. The current QA method is limited to protect the patient. The described method provides an overall analysis for dosimetric data issued from calculation and measurement and it can be quickly integrated in QA system for IMRT.

Keywords


IMRT, Quality Assurance, Statistical Method

Full Text:

PDF

References


Mijheer B, Georg D. Guidelines for the verification of IMRT. Brussels, Belgium: ESTRO; 2008.

Ezzell GA, Burmeister JW, Dogan N, LoSasso TJ, Mechalakos JG, Mihailidis D, Molineu A, Palta JR, Ramsey CR, Salter BJ, Shi J, Xia P, Yue NJ, Xiao Y: IMRT commissioning: Multiple institution planning and dosimetry comparisons, a report from AAPM Task Group 119. Med Phys 2009, 36:5359-5373.

Thwaites D. Accuracy required and achievable in radiotherapy dosimetry: have modern technology and techniques changed our views?. Journal of Physics: Conference Series 444 (2013) 012006.

Gary A E et al. IMRT commissioning: Multiple institution planning and dosimetry comparisons, a report from AAPM Task Group 119. Med. Phys 2009 ; 36 (11) : 5359-5373.

Low DA, Harms WB, Mutic S, Purdy JA. A technique for the quantitative evaluation of dose distributions. Med Phys 1998; 25(5): 656–661.

Spezi E, Lewis DG. γ histograms for radiotherapy plan evaluation. Radiotherapy and Oncology 2006; 79: 224–230.

Chaikh A, Balosso J. Should the dose prescription be readjusted when using tissues density corrections algorithms for radiation oncology?. J Case Rep Onc Ther 2014; 1(1):01018.

Chaikh A, Giraud JY, Perrin E, Bresciani J P, Balosso J. The choice of statistical methods for comparisons of dosimetric data in radiotherapy. Radiation oncology 2014; 9:205.

The R project for statistical computing. http://www.r-project.org.

Bland J Martin, Altman DG. Statistical methods for assessing agreement between two methods of clinical measurement. Lancet ,1986; i:307-310. https://www-users.york.ac.uk/~mb55/meas/ba.pdf.

Hamilton C, Stamey J. (2007). Using Bland-Altman to assess agreement between two medical devices -don’t forget the confidence intervals!. J Clin Monit Comput 21:331-33.

Mantha S, Roizen MF, Fleischer LA, et al. (2000). Comparing methods of clinical measurement: reporting standards for Bland and Altman Analysis Anesth Analg 90:593-602.

Faul F, Erdfelder E, Lang AG, Buchner A. G*Power 3: A flexible statistical power analysis program for the social, behavioral, and biomedical sciences. Behavior Research Methods 2007, 39: 175-191.

Faul F, Erdfelder E, Buchner A, Lang A G. Statistical power analyses using G*Power 3.1. Tests for correlation and regression analyses. Behavior Research Methods 2009, 41: 1149-1160.

Chaikh A, Giraud JY, Marguet M, Silva DA, Perrin E, Balosso J. A decision tool to adjust the prescribed dose after change in the dose calculation algorithm. Int J Cancer Ther Oncol 2014; 2:020414.

Zhen H. Moving from gamma passing rates to patient DVH-based QA metrics in pretreatment dose QA. Med. Phys 2011; 38 (10): 5477- 5489.

Nelms B E, Zhen H, Tomé WA. Per-beam, planar IMRT QA passing rates do not predict clinically relevant patient dose errors. Med. Phys 2011; 38, 1037–1044.

Statistical variability and confidence intervals for planar dose QA pass rates. Daniel W b, Benjamin E N, Kristopher A, Lalith K, Matthew B P. Med. Phys. 2011; 38 (11): 6053- 6064.

Stasi M, Bresciani S, Miranti A, Maggio A, Sapino V. Pretreatment patient-specific IMRT quality assurance: A correlation study between gamma index and patient clinical dose volume histogram. Med. Phys 2012; 39 (12): 7627- 7634.




DOI: http://dx.doi.org/10.14319/ijcto.34.11

Creative Commons License
This work is licensed under a Creative Commons Attribution 3.0 License.

------------------------------------------------------------

International Journal of Cancer Therapy and Oncology (ISSN 2330-4049)

© International Journal of Cancer Therapy and Oncology (IJCTO)

To make sure that you can receive messages from us, please add the 'ijcto.org' domain to your e-mail 'safe list'. If you do not receive e-mail in your 'inbox', check your 'bulk mail' or 'junk mail' folders.

------------------------------------------------------------

Number of visits since October, 2013
AmazingCounters.com