A preliminary investigation on long-term consistency of MPC as a quick daily QA application

CP Bhatt, Manoj Kumar Semwal, Sukhvir Singh, Kundan Singh Chufal, Kiran Sharma

Abstract


Purpose: The purpose of this study was to establish Machine performance check (MPC) application as a comprehensive daily QA program in a clinical setting for a True Beam 2.0 system and investigate the first ten months (195 days) daily QA data generated by the MPC.

Methods: An automated daily quality assurance (QA) application named machine performance check (MPC) was recently launched by Varian Medical Systems with their TrueBeam 2.0 linear accelerator (linac) system. MPC performs all the essential machine tests such as Beam Constancy Check, and Geometry Check with the use of an IsoCal phantom. There is no systematic published study on long-term consistency and validation of MPC in a clinical set-up for its acceptance as an alternative QA application. In the present study, we collected data with the MPC for over ten months (195 days) on a TrueBeam 2.0 system. The data was analysed for reproducibility and also compared with the data collected with other statndard QA devices at the time of commissioning of the TrueBeam system for validation.

Results: The results showed that the reproducibility of MPC was at least an order of magnitude less than the tolerance values for the respective parameters and also the average measured values for all QA parameters studied. The MPC measured isocenter accuracy, and output values were close to the Winston-Lutz test (within 0.1 mm) and the ion-chamber measurements (within 0.1%), respectively.

Conclusion: With our long term result, it is evident that the MPC could be an alternative daily QA tool. A comprehensive and long-term validation of the MPC measured values with the other standard QA methods over the ten month period will be needed before accepting MPC as a reliable QA tool.


Keywords


Radiotherapy, Machine Performance Check (MPC), TrueBeam 2.0, Linac QA

Full Text:

PDF

References


Klein EE, Hanley J, Bayouth J, et al. Quality assurance of medical accelerators: report of American association of physicists in medicine task group 142. Med. Phys. 2009;36(9):4197-212.

Bissonnette JP, Balter P, Dong L, et al. Quality assurance for image guided radiation therapy utilizing CT based technologies: a report of American association of physicists in medicine task group 179. Med Phys. 2012;39 (4):1946-63.

Clivio A, Vanetti E, Rose S, et al. Evaluation of the machine performance check application for TrueBeam Linac. Radiat Oncol. 2015;10:97.

Varian Medical Systems. TrueBeam Technical Reference Guide- manual, Palo Alto, CA. Varian Med Sys. 2013;1.

Varian Medical Systems. TrueBeam Technical Reference Guide- imaging manual, Palo Alto, CA. Varian Med Sys. 2013;2.

Gao S, Du W, Balter P, et al. Evaluation of IsoCal geometric calibration system for Varian Linacs equipped with on-board imager and electronic portal imaging device imaging systems. J Appl Clin Med Phys. 2014;15:164-81.

Ling CC, Zhang P, Archambault Y, et al. Commissioning and quality assurance of RapidArc radiotherapy delivery system. Int J Radiat Oncol Biol Phys. 2008;72:575-81.

Bedford JL, Warrington AP. Commissioning of volumetric modulated arc therapy (VMAT). Int J Radiat Oncol Biol Phys. 2009;73:537-45.

Glide-Hurst C, Bellon M, Foster R, et al. Commissioning of the Varian TrueBeam linear accelerator: a multi-institutional study. Med. Phys. 2013;40:031719.

Fontenot JD, Alkhatib H, Garrett JA, et al. AAPM medical physics practice guideline 2.a: commissioning and quality assurance of X-ray based image guided radiotherapy systems. J Appl Clin Med Phy. 2014;15(1):4528.

Rowshanfarzad P, McGarry CK, Barnes MP, et al. An EPID-based method for comprehensive verification of gantry, EPID and the MLC carriage positional accuracy in Varian linacs during arc treatments. Radiat Oncol. 2014;9:249.

Van Elmpt W, McDermott L, Nijsten S, et al. A literature review of electronic portal imaging for radiotherapy dosimetry. Radiother. Oncol. 2008;88(3):289-309.

Winkler P, Hefner A, Georg D. Dose-response characteristics of amorphous silicon EPID. Med Phys. 2005;32(10):3095-105.

Rowshanfarzad P, McGarry CK, Barnes MP, et al. An EPID-based method for comprehensive verification of gantry, EPID and the MLC carriage positional accuracy in Varian linacs during arc treatments. Radiat Oncol. 2014;9:249.

Sanghangthum T, Suriyapee S, Srisatit S, et al. Retrospective analysis of linear accelerator output constancy checks using process control techniques. J Appl Clin Med Phys. 2013;14(1):147-60.

International Atomic Energy Agency. Absorbed dose determination in external beam radiotherapy: an international code of practice for dosimetry based on standards of absorbed dose to water. IAEA. 2000.




DOI: http://dx.doi.org/10.14319/ijcto.51.17

Creative Commons License
This work is licensed under a Creative Commons Attribution 3.0 License.

------------------------------------------------------------

International Journal of Cancer Therapy and Oncology (ISSN 2330-4049)

© International Journal of Cancer Therapy and Oncology (IJCTO)

To make sure that you can receive messages from us, please add the 'ijcto.org' domain to your e-mail 'safe list'. If you do not receive e-mail in your 'inbox', check your 'bulk mail' or 'junk mail' folders.

------------------------------------------------------------

Number of visits since October, 2013
AmazingCounters.com